Model
|
AWA6291
|
Name
|
Real Time Signal Analyzer
|
Fulfils Standards
|
IEC 61672 class 1 and IEC 61260 class 1
|
Microphone
|
1/2" Prepolarized Condenser Mic. Model AWA14423 (sensitivity: 50 mV/Pa)
|
Preamplifier
|
AWA14601
|
Frequency Range
|
10 Hz ~ 20 kHz
|
Range Gain
|
-10 dB, 0 dB, 10 dB, 20 dB, 30 dB, 40 dB.
Range Error: ≤0.1 dB.
|
Self-generated Noise
|
<18 dB(A), 23 dB(C), 28 dB(Z)
|
Meas. upper Range
|
140 dB
|
Peak C Sound Level Measuring Range
|
50~143 dB
|
Frequency Weighting.
|
Parallel (simultaneous) A, C, Z. Realized by digital filtering
|
Time Weighting
|
Parallel F, S, I, Peak C+, Peak C-
|
A/D Bits
|
24 bits
|
Level Linearity Range
|
>90 dB
|
Sampling Frequency
|
48 kHz
|
Display
|
240×160 graphic LCD, with EL back-lit
|
Display Refresh
|
1 Hz for value; 10 Hz for graph
|
Main Parameters of Integrating Measurement
|
Lxyp, Lxyi, Lxeq,1s, Lxeq,T, LAE, E, Cpeak+, Cpeak-, LAFmax, LAFmin, LAFeqT, LASeqT, LAIeqT
Note: X is A, C, Z and Y is F, S, I
|
Main Parameters of Airport Noise Meas
|
LEPN,Td,Lmax
|
Display Resolution
|
0.1 dB
|
Real-time Analysis Speed
|
Approx. 47 times/s (It can simultaneously complete all center frequencies and A, C, Z weighting.)
|
Statistical analysis Function
|
Statistical analysis software (S6291-00101)
Statistical analysis Menu: LAFp, LAFmax,
LAF5, LAF10, LAF50, LAF90, LAF95, LAFmin, SD, LAeq1 s, LAeq, T, LAE, LAFeqT.
24 hours monitors menu :And Ld, Ln, Ldn
|
Real-time spectrum analysis
|
Fulfils standards: IEC 61260 Class 1
Filter type: Digital band-pass filter
Measuring Parameters: Lfmi, Lfmeq,1s, Lfmeq,T, Lxyp, Lxyi, Lxeq,1s, Lxeq,T, LAE, E, Cpeak, Tm
Note: fm is center frequency; X is frequency weighting A, C, Z. Y is time weighting F, S, I
|
Real-time 1/1 OCT spectrum analysis
|
1/1 OCT spectrum analysis software for noise (S6291 -00202)
Filters Center Frequency: 16 Hz~16 kHz
|
Real-time 1/1 OCT spectrum analysis
|
1/3 OCT spectrum analysis software for noise (S6291 -00303)
Filters Center Frequency: 10 Hz~20 kHz
|
FFT analysis function
|
FFT analysis software for noise (S6291 -00404)
1. Line Number: 200 lines on the display
2. Sampling Frequency: 48 kHz, 24 kHz, 12 kHz,6 kHz, 3 kHz, 1.5 kHz.
3. Upper Limit of Analysis Frequency: 18.75 kHz, 9375 Hz, 4687 Hz, 2344 Hz, 1172 Hz, 586 Hz
4. Bandwidth of each line: 93.75 Hz, 46.875 Hz, 23.438 Hz, 11.74 Hz, 5.8594 Hz, 2.9297 Hz
5. Average Method: index and linearity
6. Measuring Parameters: Lxyp, Lxyi, Lxeq,1s, Lxeq,T, LAE, E, Cpeak,Tm in every line
|
Reverberation time measurement function
|
Reverberation time measurement software (S6291-00508)
1. Function: Reverberation time, EDT measurement by integrated impulse response method.
2. Fulfils Standard: ISO 3382
3. Measuring Parameters: T20, T30, EDT
4. Measuring Range: 0.20s to 20.00 s
5. Wave record time: 10 s or 20 s
6. Bandwidth of digital filters:1/1 OCT or 1/3 OCT
7. Center Frequency Ranges: 63 Hz ~ 16 kHz for Oct, 40 Hz ~ 16 kHz for 1/3 Oct
8. Date store: Max 12288 groups reverberation time, also translate to U disc
|
Data Storage
|
64M bits Flash, instantaneous data, 1 second integral results and integral measuring result (maximum 12288 groups). Also can be stored in Flash disk or transferred to the Flash disk through USB interface
|
Integral Time
|
Manual, 10 s, 1 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1h, 4h, 8h, 24h or 1 min~59 min
|
Print
|
Can print analytical results in the mini-print, can also print the contents showed on LCD
|
Data Recall
|
Recalled any data
|
Intelligent Delete
|
Burst event may be deleted scientifically
|
Internal Clock
|
Error less than 1 min/month
|
Output Interface
|
Stereo Connector AC Signal Output
RS 232 interface to the mini-printer, or connected to PC
USB Interface, which can transfer the measuring results to PC, or the Flash disk
|
Power Supply
|
6×LR6 alkaline battery or rechargeable batteries, which can work for over 12 hours continuously. 9 V external power supply is also acceptable
|
Dimensions
|
310(H) × 90(W) × 38(D),mm
|
Weight
|
0.5 kg (include batteries)
|
EMC
|
Type X
|
Working Temperature
|
-10 ~ 50°c
|
Relative Humidity
|
25 ~ 90%
|
Atmospheric Pressure
|
65 kPa ~ 108 kPa
|