Additel
Refine Search
Additel 760 Portable Handheld Automatic Pressure Calibrators
Automatic Handheld Pressure Calibrator
|
Overview
A portable automated pressure calibrator in the palm of your hand—this could be our most exciting product yet! The Additel 760 series Automatic Handheld Pressure Calibrator takes portable pressure calibration to new levels. Weighing less than 4 lbs (1.8 kg), the ADT760’s innovative design contains a built-in pump, precision pressure sensor, internal controller and a large touch-screen colour display. To generate pressure, simply key in the desired pressure and the Additel 760 will do the rest. Each unit has four channels: one internal pressure channel for source and measure pressures, two external pressure measurement channels, and one electronic measure and source channel. This series of calibrator has three standard models with the option of adding HART communications, documentation and data logging. |
Documenting Process Functionality Each model of the Additel 760 series has an option incorporating documentation and communication functions turning your 760 into a multifunction documenting process calibrator. This feature provides HART communication, task documentation and data logging. |
ADT760-LLP The 760-LLP is designed for low pressure calibration and comes with a built-in pressure module to ±30 inH2O (±75 mbar) and provides an accuracy to 0.05%FS (see ordering information for configurations with the option of the ADT760 and a module of your choice). Additional internal pressure modules (ADT155 series) are available and provide a variety of ranges down to ±0.25 inH2O (±0.62 mbar). The accuracy of 0.05%FS and control stability 0.005%FS is based on the internal module’s span. Measurements can be made in gauge or differential mode. |
ADT760-D The 760-D gives you differential and gauge pressure but at a higher-pressure range than the ADT760-LLP. Covering the range of -12.5 to 35 psi (-0.86 to 2.5 bar) and with an accuracy of 0.02%FS, the ADT760-D is an ideal solution to cover very common gauge and differential pressure measurements. The Additel 760-D comes with an internal module covering its full range, but lower pressure configurations down to ±10 inH2O differential can be purchased to improve accuracy at lower pressures. |
ADT760-MA The 760-MA generates and controls pressure from vacuum pressures up to 300 psig (20 bar.g) with an accuracy of 0.02%FS. Equipped with a built-in barometric reference, each unit can switch between gauge and absolute pressure types. A variety of internal sensors are available which offer lower pressure ranges for improved performance. |
Applications
- Gauge calibration for differential and gauge type pressures
- Transmitter calibration for absolute and gauge type pressures
- Switch calibration for absolute and gauge type pressures
- HART-smart transmitter calibration (-DL version only)
- Pressure monitoring and data logging (-DL version only)
|
|
|
|
Specifications
|
|
|
Order Info
|
Additel 761A Automated Pressure Calibrator
Automated Pressure Calibrator
|
Overview
At Additel, innovation and continuous improvement are part of our company's culture and the products we introduce. When we set out to deliver the Additel 761A series calibrators, we knew we needed to provide breakthrough improvements and additional value to the existing line of calibrators (Additel 761 series). The ADT761A has many improvements: increased pressure range to 1,000 psi (70 bar), removable internal pressure modules, optional precision models to 0.01%FS, increased speed to pressure, ability to read two external pressure modules, touch screen display, Wi-Fi, Bluetooth, and Ethernet communications, double the original battery life, and more! |
ADT761A-LLP The Additel 761A-LLP is designed for low pressure calibration and comes with a ±30 inH2O (±75 mbar) high range module and a low range module of your choice ranging from ±20 inH2O to as low as ±0.25 inH2O (±50 to ±0.62 mbar). This unit has an accuracy of 0.05%FS with control stability better than 0.005%FS. All measurements can be made in differential or gauge pressures. |
ADT761A-D The Additel 761A-D also provides differential and gauge measurement which covers the range of -13.5 to 35 psi (-0.95 to 2.5 bar). This unit comes with a CP35 module (-13.5 to 35 psi) and one low range module of your choice ranging as low as ±10 inH2O (25 mbar). Each ADT761A-D can be preconfigured with the modules that fit your need to give you the best precision at the pressures you perform calibrations. |
ADT761A-500 The Additel 761A-500 will generate and control from vacuum pressures up to 500 psig (35 bar.g). Both gauge and absolute pressures can be realized due to a built-in barometer. Each unit comes with a CP500 module (-12.5 to 500 psig) for the high range and the low range can be preconfigured based on the variety of modules available down to 10 psig (0.7 bar.g). |
ADT761A-1K The Additel 761A-1K will quickly generate and control from vacuum pressures up to 1,000 psig (70 bar.g). Like the ADT761A-500, both gauge and absolute pressures can be realized due to a built-in barometer. Each unit comes with a CP1K module (-12.5 to 1,000 psig) for the high range and the low range can be preconfigured based on the variety of modules available down to 30 psig (2 bar.g). |
ADT761A-BP The Additel 761A-BP is designed for calibration of barometer sensors. With a range of 100 to 1200 hPa and an accuracy of 0.01%FS, this unit is ideal for calibration on the bench or in the field. |
ADT761A-APXR Precision Accuracy Options The Additel 761A series includes a precision accuracy option which provides an accuracy of 0.01%FS. This calibrator option includes a single non-removable sensor and can measure absolute and gauge pressures. Model configurations are available from 15 to 1,000 psig (1 to 70 bar.g). |
Applications
High Pressure Automatic Calibration
|
Pressure Gauge, Transmitter, or Switch Calibration
|
Task Menu
|
Specifications
|
|
|
|
ADT875-155
Dry Well Calibrator
|
Overview
If you are serious about portable temperature calibration tools, then you know a good dry well calibrator is more than just a stable heat source. The Additel 875 Series Dry Well Calibrators combine excellent performance in stability, radial and axial uniformity. And loading with speed, ruggedness and portability. But we don’t stop there! The Process Calibrator option adds the capabilities of a three-channel thermometer readout and a documenting process calibrator. We’ve also incorporated a unique option to select your own temperature range within the range of the model selected. We’re calling this the CYOR option or Choose Your Own Range option. When you purchase the CYOR option, you pick the upper and lower temperature range needed and we calibrate and optimize the dry well’s performance over your selected range. Each unit has a colour touch screen display, dual-zone control, and much more. You are just going to love these new dry wells! |
Specifications
Specification |
875-155 |
875-350 |
875-660 |
||
Temperature Range at 23℃ |
-40°C to 155°C |
33°C to 350°C |
33°C to 660°C |
||
Display Accuracy |
±0.18°C at Full Range |
±0.2°C at Full Range |
±0.3°C at 33°C ±0.3°C at 420°C ±0.5°C at 660°C |
||
Stability(30 min) |
±0.01°C at Full Range |
±0.02°C at Full Range |
±0.02°C at 33°C ±0.03°C at 50°C ±0.04°C at 420°C ±0.04°C at 660°C |
||
Axial Uniformity at 60 mm(2.4 in) |
±0.07°C at Full Range |
±0.04°C at 33°C ±0.1°C at 200°C ±0.2°C at 350°C |
±0.05°C at 33°C ±0.3°C at 420°C ±0.5°C at 660°C |
||
Radial Uniformity |
±0.01°C at Full Range |
±0.01°C at 33°C ±0.015°C at 200°C ±0.02°C at 350°C |
±0.02°C at 33°C ±0.05°C at 420°C ±0.1°C at 660°C |
||
Loading Effect |
±0.1°C (Display Sensor) ±0.02°C(External Sensor) |
±0.15°C(Display Sensor) ±0.015°C(ExternalSensor) |
±0.15°C(DisplaySensor) ±0.025°C (External Sensor) |
||
Hysteresis (Display Sensor) |
0.025°C |
0.03°C |
0.1°C |
||
Environmental Conditions |
8°C to 38°C guaranteed accuracy 0°C to 50°C, 0% to 90% RH non-condensing, 3000 M altitude for normal operation |
||||
Storage Conditions |
-20°C to 60°C |
||||
IP Rating |
IP20 |
||||
Immersion Depth |
150 mm (5.9 in) |
||||
Insert OD |
25.8 mm (1.02 in) |
24.8 mm (0.98 in) |
|||
Heating Time |
13 min: -40°C to 155°C 5 min: -40°C to 23°C 8 min: 23°C to 155°C |
5 min: 33°C to 350°C |
15 min: 33°C to 660°C |
||
Cooling Time |
28 min: 155°C to -40°C 8 min: 155°C to 23°C 20 min: 23°C to -40°C |
15 min: 350°C to 100°C 10 min: 100°C to 50°C 10 min: 50°C to 33°C |
23 min: 660°C to 100°C 12 min: 100°C to 50°C 12 min: 50°C to 33°C |
||
Typical Time to Stability |
10 min |
||||
Resolution |
0.01°C |
||||
Units |
°C, °F, and K |
||||
Display |
6.5 in (165 mm) colour touch screen |
||||
Size (H x W x D) |
320 x 170 x 330 mm (12.6 x 6.7 x 13.0 in) |
||||
Weight |
9.9 kg (21.8 lbs) |
8.2 kg (18.1 lbs) |
|||
Voltage |
90-254 VAC, 45-65 Hz, 580 W |
90-254 VAC, 45-65 Hz, 1200 W |
|||
Mechanical Testing |
Vibration: 2 g (10-500 Hz), 30 min for 2 sides Impact: 4 g three times Drop test: 500 mm (19.6 in) |
||||
Communication |
USB A, USB B, RJ45, WiFi, Bluetooth |
||||
Localization |
English, Chinese, Japanese, Russian, German, French, Italian, and Spanish |
||||
Warranty |
1 year |
Specification ADT875PC
Specification |
Description |
Readout Accuracy for 100ohm PRT (Probe Accuracy Not Included) |
±0.009°C at -40°C ±0.010°C at 0°C ±0.012°C at 50°C ±0.017°C at 155°C ±0.019°C at 200°C ±0.026°C at 350°C ±0.030°C at 420°C ±0.042°C at 660°C |
Readout Resolution |
1 mΩ |
Reference Resistance Range |
0 Ω to 400 Ω |
Reference Resistance Accuracy |
0 Ω to 50 Ω: 0.002 Ω 50 Ω to 400 Ω: 0.004% RD |
Reference Characterizations |
ITS-90, CVD, IEC-751, Resistance |
Reference Measurement Capability |
4-wire PRT |
Reference Probe Connection |
6-pin lemo smart connector |
RTD Channels |
2 |
RTD Measurement Accuracy (excl sensor) Compliance |
0 Ω to 25 Ω: 0.002 Ω 25 Ω to 400 Ω: 0.004% RD 400 Ω to 4K Ω: 0.008% RD |
RTD Measurement Resolution |
0 Ω to 400 Ω: 1 mΩ 400 Ω to 4K Ω: 0.01 Ω |
RTD Measurement Resistance Range |
0 Ω to 4K Ω |
RTD Characterizations |
PT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120 |
RTD Connection |
Four 4 mm input jacks |
RTD Channels |
2 channels. Both accept 2, 3, or 4-wire RTDs |
TC Channel |
2 |
TC Measurement Channels |
Mini TC terminals: Accepting S, R, K, B, N, E, J, T, C, D, G, L, and U |
TC Measurement Accuracy (excl sensor) |
ype K: ±0.13°C at 0°C ±0.15°C at 155°C ±0.18°C at 350°C ±0.24°C at 660°C |
TC Range |
–100 mV to 100 mV |
TC Resolution |
0.001 mV, Input Impedance<1 MΩ |
TC Voltage Accuracy |
0.02% RD + 5 µV |
Internal CJC Accuracy |
±0.35°C (ambient from 0 °C to 50 °C) |
Current Range |
–30 mA to 30 mA |
Current Accuracy |
0.02% RD + 2 µA |
Current Resolution |
0.001 mA, Input Impedance: < 10Ω |
Voltage Range |
–30 V to 30 V |
Voltage Accuracy |
±0.02% RD + 2 mV |
Voltage Resolution |
0.001V; Input impedance: < 1MΩ |
Switch Test |
Mechanical or Electrical |
DC 24V Output |
24 V ±1 V, MAX60 mA |
Hart Communication |
Optional (ADT875PC Model) |
Documentation |
Up to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions. |
Temperature Coefficient 0℃ to 8℃ and 38℃ to 50℃ |
ADT875(PC)-155: ±0.005 °C/°C ADT875(PC)-350/660: ±0.01 °C/°C Ref Readout: ±1 ppm FS/°C RTD Readouts: ±2 ppm FS/°C TC Readouts: ±5 ppm FS/°C Current: ±10 ppm FS/°C Voltage: ±10 ppm FS/°C |
TC Type |
Temperature (°C) |
Error (°C)[1] |
TC Type |
Temperature (°C) |
Error (°C)[1] |
N |
-40 0 155 350 660 |
±0.2 ±0.2 ±0.19 ±0.2 ±0.24 |
L |
-40 0 155 350 660 |
±0.1 ±0.1 ±0.12 ±0.16 ±0.21 |
E |
-40 0 155 350 660 |
±0.09 ±0.09 ±0.1 ±0.13 ±0.19 |
R |
-40 0 155 350 660 |
±1.23 ±0.95 ±0.63 ±0.56 ±0.54 |
J |
-40 0 155 350 660 |
±0.1 ±0.1 ±0.12 ±0.16 ±0.21 |
T |
-40 0 155 350 400 |
±0.14 ±0.13 ±0.13 ±0.15 ±0.15 |
K |
-40 0 155 350 660 |
±0.13 ±0.13 ±0.16 ±0.19 ±0.25 |
U |
-40 0 155 350 660 |
±0.14 ±0.13 ±0.13 ±0.14 ±0.17 |
C |
0 155 350 660 |
±0.38 ±0.34 ±0.33 ±0.38 |
D |
0 155 350 660 |
±0.52 ±0.37 ±0.33 ±0.36 |
B |
250 350 660 |
±2 ±1.44 ±0.84 |
[1] Excluding cold junction compensation errors |
Applications
Order Info
|
ADT875-350 Temperature Calibrator
Additel 875-350
Dry Well Calibrator
|
Overview
If you are serious about portable temperature calibration tools, then you know a good dry well calibrator is more than just a stable heat source. The Additel 875 Series Dry Well Calibrators combine excellent performance in stability, radial and axial uniformity. And loading with speed, ruggedness and portability. But we don’t stop there! The Process Calibrator option adds the capabilities of a three-channel thermometer readout and a documenting process calibrator. We’ve also incorporated a unique option to select your own temperature range within the range of the model selected. We’re calling this the CYOR option or Choose Your Own Range option. When you purchase the CYOR option, you pick the upper and lower temperature range needed and we calibrate and optimize the dry well’s performance over your selected range. Each unit has a colour touch screen display, dual-zone control, and much more. You are just going to love these new dry wells! |
Specifications
Specification |
875-155 |
875-350 |
875-660 |
||
Temperature Range at 23℃ |
-40°C to 155°C |
33°C to 350°C |
33°C to 660°C |
||
Display Accuracy |
±0.18°C at Full Range |
±0.2°C at Full Range |
±0.3°C at 33°C ±0.3°C at 420°C ±0.5°C at 660°C |
||
Stability(30 min) |
±0.01°C at Full Range |
±0.02°C at Full Range |
±0.02°C at 33°C ±0.03°C at 50°C ±0.04°C at 420°C ±0.04°C at 660°C |
||
Axial Uniformity at 60 mm(2.4 in) |
±0.07°C at Full Range |
±0.04°C at 33°C ±0.1°C at 200°C ±0.2°C at 350°C |
±0.05°C at 33°C ±0.3°C at 420°C ±0.5°C at 660°C |
||
Radial Uniformity |
±0.01°C at Full Range |
±0.01°C at 33°C ±0.015°C at 200°C ±0.02°C at 350°C |
±0.02°C at 33°C ±0.05°C at 420°C ±0.1°C at 660°C |
||
Loading Effect |
±0.1°C (Display Sensor) ±0.02°C(External Sensor) |
±0.15°C(Display Sensor) ±0.015°C(ExternalSensor) |
±0.15°C(DisplaySensor) ±0.025°C (External Sensor) |
||
Hysteresis (Display Sensor) |
0.025°C |
0.03°C |
0.1°C |
||
Environmental Conditions |
8°C to 38°C guaranteed accuracy 0°C to 50°C, 0% to 90% RH non-condensing, 3000 M altitude for normal operation |
||||
Storage Conditions |
-20°C to 60°C |
||||
IP Rating |
IP20 |
||||
Immersion Depth |
150 mm (5.9 in) |
||||
Insert OD |
25.8 mm (1.02 in) |
24.8 mm (0.98 in) |
|||
Heating Time |
13 min: -40°C to 155°C 5 min: -40°C to 23°C 8 min: 23°C to 155°C |
5 min: 33°C to 350°C |
15 min: 33°C to 660°C |
||
Cooling Time |
28 min: 155°C to -40°C 8 min: 155°C to 23°C 20 min: 23°C to -40°C |
15 min: 350°C to 100°C 10 min: 100°C to 50°C 10 min: 50°C to 33°C |
23 min: 660°C to 100°C 12 min: 100°C to 50°C 12 min: 50°C to 33°C |
||
Typical Time to Stability |
10 min |
||||
Resolution |
0.01°C |
||||
Units |
°C, °F, and K |
||||
Display |
6.5 in (165 mm) colour touch screen |
||||
Size (H x W x D) |
320 x 170 x 330 mm (12.6 x 6.7 x 13.0 in) |
||||
Weight |
9.9 kg (21.8 lbs) |
8.2 kg (18.1 lbs) |
|||
Voltage |
90-254 VAC, 45-65 Hz, 580 W |
90-254 VAC, 45-65 Hz, 1200 W |
|||
Mechanical Testing |
Vibration: 2 g (10-500 Hz), 30 min for 2 sides Impact: 4 g three times Drop test: 500 mm (19.6 in) |
||||
Communication |
USB A, USB B, RJ45, WiFi, Bluetooth |
||||
Localization |
English, Chinese, Japanese, Russian, German, French, Italian, and Spanish |
||||
Warranty |
1 year |
Specification ADT875PC
Specification |
Description |
Readout Accuracy for 100ohm PRT (Probe Accuracy Not Included) |
±0.009°C at -40°C ±0.010°C at 0°C ±0.012°C at 50°C ±0.017°C at 155°C ±0.019°C at 200°C ±0.026°C at 350°C ±0.030°C at 420°C ±0.042°C at 660°C |
Readout Resolution |
1 mΩ |
Reference Resistance Range |
0 Ω to 400 Ω |
Reference Resistance Accuracy |
0 Ω to 50 Ω: 0.002 Ω 50 Ω to 400 Ω: 0.004% RD |
Reference Characterizations |
ITS-90, CVD, IEC-751, Resistance |
Reference Measurement Capability |
4-wire PRT |
Reference Probe Connection |
6-pin lemo smart connector |
RTD Channels |
2 |
RTD Measurement Accuracy (excl sensor) Compliance |
0 Ω to 25 Ω: 0.002 Ω 25 Ω to 400 Ω: 0.004% RD 400 Ω to 4K Ω: 0.008% RD |
RTD Measurement Resolution |
0 Ω to 400 Ω: 1 mΩ 400 Ω to 4K Ω: 0.01 Ω |
RTD Measurement Resistance Range |
0 Ω to 4K Ω |
RTD Characterizations |
PT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120 |
RTD Connection |
Four 4 mm input jacks |
RTD Channels |
2 channels. Both accept 2, 3, or 4-wire RTDs |
TC Channel |
2 |
TC Measurement Channels |
Mini TC terminals: Accepting S, R, K, B, N, E, J, T, C, D, G, L, and U |
TC Measurement Accuracy (excl sensor) |
ype K: ±0.13°C at 0°C ±0.15°C at 155°C ±0.18°C at 350°C ±0.24°C at 660°C |
TC Range |
–100 mV to 100 mV |
TC Resolution |
0.001 mV, Input Impedance<1 MΩ |
TC Voltage Accuracy |
0.02% RD + 5 µV |
Internal CJC Accuracy |
±0.35°C (ambient from 0 °C to 50 °C) |
Current Range |
–30 mA to 30 mA |
Current Accuracy |
0.02% RD + 2 µA |
Current Resolution |
0.001 mA, Input Impedance: < 10Ω |
Voltage Range |
–30 V to 30 V |
Voltage Accuracy |
±0.02% RD + 2 mV |
Voltage Resolution |
0.001V; Input impedance: < 1MΩ |
Switch Test |
Mechanical or Electrical |
DC 24V Output |
24 V ±1 V, MAX60 mA |
Hart Communication |
Optional (ADT875PC Model) |
Documentation |
Up to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions. |
Temperature Coefficient 0℃ to 8℃ and 38℃ to 50℃ |
ADT875(PC)-155: ±0.005 °C/°C ADT875(PC)-350/660: ±0.01 °C/°C Ref Readout: ±1 ppm FS/°C RTD Readouts: ±2 ppm FS/°C TC Readouts: ±5 ppm FS/°C Current: ±10 ppm FS/°C Voltage: ±10 ppm FS/°C |
TC Type |
Temperature (°C) |
Error (°C)[1] |
TC Type |
Temperature (°C) |
Error (°C)[1] |
N |
-40 0 155 350 660 |
±0.2 ±0.2 ±0.19 ±0.2 ±0.24 |
L |
-40 0 155 350 660 |
±0.1 ±0.1 ±0.12 ±0.16 ±0.21 |
E |
-40 0 155 350 660 |
±0.09 ±0.09 ±0.1 ±0.13 ±0.19 |
R |
-40 0 155 350 660 |
±1.23 ±0.95 ±0.63 ±0.56 ±0.54 |
J |
-40 0 155 350 660 |
±0.1 ±0.1 ±0.12 ±0.16 ±0.21 |
T |
-40 0 155 350 400 |
±0.14 ±0.13 ±0.13 ±0.15 ±0.15 |
K |
-40 0 155 350 660 |
±0.13 ±0.13 ±0.16 ±0.19 ±0.25 |
U |
-40 0 155 350 660 |
±0.14 ±0.13 ±0.13 ±0.14 ±0.17 |
C |
0 155 350 660 |
±0.38 ±0.34 ±0.33 ±0.38 |
D |
0 155 350 660 |
±0.52 ±0.37 ±0.33 ±0.36 |
B |
250 350 660 |
±2 ±1.44 ±0.84 |
[1] Excluding cold junction compensation errors |
Applications
Order Info
|
ADT875-660
Dry Well Calibrator
|
Overview
If you are serious about portable temperature calibration tools, then you know a good dry well calibrator is more than just a stable heat source. The Additel 875 Series Dry Well Calibrators combine excellent performance in stability, radial and axial uniformity. And loading with speed, ruggedness and portability. But we don’t stop there! The Process Calibrator option adds the capabilities of a three-channel thermometer readout and a documenting process calibrator. We’ve also incorporated a unique option to select your own temperature range within the range of the model selected. We’re calling this the CYOR option or Choose Your Own Range option. When you purchase the CYOR option, you pick the upper and lower temperature range needed and we calibrate and optimize the dry well’s performance over your selected range. Each unit has a colour touch screen display, dual-zone control, and much more. You are just going to love these new dry wells! |
Specifications
Specification |
875-155 |
875-350 |
875-660 |
||
Temperature Range at 23℃ |
-40°C to 155°C |
33°C to 350°C |
33°C to 660°C |
||
Display Accuracy |
±0.18°C at Full Range |
±0.2°C at Full Range |
±0.3°C at 33°C ±0.3°C at 420°C ±0.5°C at 660°C |
||
Stability(30 min) |
±0.01°C at Full Range |
±0.02°C at Full Range |
±0.02°C at 33°C ±0.03°C at 50°C ±0.04°C at 420°C ±0.04°C at 660°C |
||
Axial Uniformity at 60 mm(2.4 in) |
±0.07°C at Full Range |
±0.04°C at 33°C ±0.1°C at 200°C ±0.2°C at 350°C |
±0.05°C at 33°C ±0.3°C at 420°C ±0.5°C at 660°C |
||
Radial Uniformity |
±0.01°C at Full Range |
±0.01°C at 33°C ±0.015°C at 200°C ±0.02°C at 350°C |
±0.02°C at 33°C ±0.05°C at 420°C ±0.1°C at 660°C |
||
Loading Effect |
±0.1°C (Display Sensor) ±0.02°C(External Sensor) |
±0.15°C(Display Sensor) ±0.015°C(ExternalSensor) |
±0.15°C(DisplaySensor) ±0.025°C (External Sensor) |
||
Hysteresis (Display Sensor) |
0.025°C |
0.03°C |
0.1°C |
||
Environmental Conditions |
8°C to 38°C guaranteed accuracy 0°C to 50°C, 0% to 90% RH non-condensing, 3000 M altitude for normal operation |
||||
Storage Conditions |
-20°C to 60°C |
||||
IP Rating |
IP20 |
||||
Immersion Depth |
150 mm (5.9 in) |
||||
Insert OD |
25.8 mm (1.02 in) |
24.8 mm (0.98 in) |
|||
Heating Time |
13 min: -40°C to 155°C 5 min: -40°C to 23°C 8 min: 23°C to 155°C |
5 min: 33°C to 350°C |
15 min: 33°C to 660°C |
||
Cooling Time |
28 min: 155°C to -40°C 8 min: 155°C to 23°C 20 min: 23°C to -40°C |
15 min: 350°C to 100°C 10 min: 100°C to 50°C 10 min: 50°C to 33°C |
23 min: 660°C to 100°C 12 min: 100°C to 50°C 12 min: 50°C to 33°C |
||
Typical Time to Stability |
10 min |
||||
Resolution |
0.01°C |
||||
Units |
°C, °F, and K |
||||
Display |
6.5 in (165 mm) colour touch screen |
||||
Size (H x W x D) |
320 x 170 x 330 mm (12.6 x 6.7 x 13.0 in) |
||||
Weight |
9.9 kg (21.8 lbs) |
8.2 kg (18.1 lbs) |
|||
Voltage |
90-254 VAC, 45-65 Hz, 580 W |
90-254 VAC, 45-65 Hz, 1200 W |
|||
Mechanical Testing |
Vibration: 2 g (10-500 Hz), 30 min for 2 sides Impact: 4 g three times Drop test: 500 mm (19.6 in) |
||||
Communication |
USB A, USB B, RJ45, WiFi, Bluetooth |
||||
Localization |
English, Chinese, Japanese, Russian, German, French, Italian, and Spanish |
||||
Warranty |
1 year |
Specification ADT875PC
Specification |
Description |
Readout Accuracy for 100ohm PRT (Probe Accuracy Not Included) |
±0.009°C at -40°C ±0.010°C at 0°C ±0.012°C at 50°C ±0.017°C at 155°C ±0.019°C at 200°C ±0.026°C at 350°C ±0.030°C at 420°C ±0.042°C at 660°C |
Readout Resolution |
1 mΩ |
Reference Resistance Range |
0 Ω to 400 Ω |
Reference Resistance Accuracy |
0 Ω to 50 Ω: 0.002 Ω 50 Ω to 400 Ω: 0.004% RD |
Reference Characterizations |
ITS-90, CVD, IEC-751, Resistance |
Reference Measurement Capability |
4-wire PRT |
Reference Probe Connection |
6-pin lemo smart connector |
RTD Channels |
2 |
RTD Measurement Accuracy (excl sensor) Compliance |
0 Ω to 25 Ω: 0.002 Ω 25 Ω to 400 Ω: 0.004% RD 400 Ω to 4K Ω: 0.008% RD |
RTD Measurement Resolution |
0 Ω to 400 Ω: 1 mΩ 400 Ω to 4K Ω: 0.01 Ω |
RTD Measurement Resistance Range |
0 Ω to 4K Ω |
RTD Characterizations |
PT10, PT25, PT50, PT100, PT200, PT500, PT1000, CU10, CU50, CU100, NI100, NI120 |
RTD Connection |
Four 4 mm input jacks |
RTD Channels |
2 channels. Both accept 2, 3, or 4-wire RTDs |
TC Channel |
2 |
TC Measurement Channels |
Mini TC terminals: Accepting S, R, K, B, N, E, J, T, C, D, G, L, and U |
TC Measurement Accuracy (excl sensor) |
ype K: ±0.13°C at 0°C ±0.15°C at 155°C ±0.18°C at 350°C ±0.24°C at 660°C |
TC Range |
–100 mV to 100 mV |
TC Resolution |
0.001 mV, Input Impedance<1 MΩ |
TC Voltage Accuracy |
0.02% RD + 5 µV |
Internal CJC Accuracy |
±0.35°C (ambient from 0 °C to 50 °C) |
Current Range |
–30 mA to 30 mA |
Current Accuracy |
0.02% RD + 2 µA |
Current Resolution |
0.001 mA, Input Impedance: < 10Ω |
Voltage Range |
–30 V to 30 V |
Voltage Accuracy |
±0.02% RD + 2 mV |
Voltage Resolution |
0.001V; Input impedance: < 1MΩ |
Switch Test |
Mechanical or Electrical |
DC 24V Output |
24 V ±1 V, MAX60 mA |
Hart Communication |
Optional (ADT875PC Model) |
Documentation |
Up to 1,000 tasks which store up to 10 results each containing as found and as left data. Snap shot feature allows for screen captures. Records auto step and ramp functions. |
Temperature Coefficient 0℃ to 8℃ and 38℃ to 50℃ |
ADT875(PC)-155: ±0.005 °C/°C ADT875(PC)-350/660: ±0.01 °C/°C Ref Readout: ±1 ppm FS/°C RTD Readouts: ±2 ppm FS/°C TC Readouts: ±5 ppm FS/°C Current: ±10 ppm FS/°C Voltage: ±10 ppm FS/°C |
TC Type |
Temperature (°C) |
Error (°C)[1] |
TC Type |
Temperature (°C) |
Error (°C)[1] |
N |
-40 0 155 350 660 |
±0.2 ±0.2 ±0.19 ±0.2 ±0.24 |
L |
-40 0 155 350 660 |
±0.1 ±0.1 ±0.12 ±0.16 ±0.21 |
E |
-40 0 155 350 660 |
±0.09 ±0.09 ±0.1 ±0.13 ±0.19 |
R |
-40 0 155 350 660 |
±1.23 ±0.95 ±0.63 ±0.56 ±0.54 |
J |
-40 0 155 350 660 |
±0.1 ±0.1 ±0.12 ±0.16 ±0.21 |
T |
-40 0 155 350 400 |
±0.14 ±0.13 ±0.13 ±0.15 ±0.15 |
K |
-40 0 155 350 660 |
±0.13 ±0.13 ±0.16 ±0.19 ±0.25 |
U |
-40 0 155 350 660 |
±0.14 ±0.13 ±0.13 ±0.14 ±0.17 |
C |
0 155 350 660 |
±0.38 ±0.34 ±0.33 ±0.38 |
D |
0 155 350 660 |
±0.52 ±0.37 ±0.33 ±0.36 |
B |
250 350 660 |
±2 ±1.44 ±0.84 |
[1] Excluding cold junction compensation errors |
Applications
Order Info
|
Additel 286
Additel 286
|
|
Applications |
Hydraulic Pressure Test Pumps
Additel 928
|
|
Additel 959A
|
Additel 925
|
|
Additel 927
|
|
Additel 936
|
|
Additel 937
|
|
Additel 938
|
|
Additel 946A
|
|
Additel 949
|
|
Additel 960
|
Pneumatic Test Pumps
Additel 901A
|
|
Additel 914A
|
|
Additel 916
|
|
Additel 912A
|
|
Additel 917
|
|
Additel 918
|
|
Additel 919A
|
|
Additel 920
|
|
Additel 920HV
|
|
ADDITEL ACCESSORIES and SOFTWARE
Additel 102
|
|
Additel 103
|
|
Additel 104-HP
|
|
Additel 100
|
|
Additel 12X
|
|
Additel 100 Series Filter
|
|
ACal
• Supports multi-users and network environments • Calibration and asset management • Simple user interface • Scanning and printing of QR codes • User definable permissions and access levels • Preset test configurations • Can calibrate variety pressure instruments • Can calibrate several instruments at a time • Calibration planning and scheduling • Certificate management and creation • Certificate customization |
ADDITEL AUTOMATED PRESSURE CALIBRATORS
Additel 761A
Automated Pressure Calibrator
|
|
Overview
At Additel, innovation and continuous improvement are part of our company's culture and the products we introduce. When we set out to deliver the Additel 761A series calibrators, we knew we needed to provide breakthrough improvements and additional value to the existing line of calibrators (Additel 761 series). The ADT761A has many improvements: increased pressure range to 1,000 psi (70 bar), removable internal pressure modules, optional precision models to 0.01%FS, increased speed to pressure, ability to read two external pressure modules, touch screen display, Wi-Fi, Bluetooth, and Ethernet communications, double the original battery life, and more! |
ADT761A-LLP The Additel 761A-LLP is designed for low pressure calibration and comes with a ±30 inH2O (±75 mbar) high range module and a low range module of your choice ranging from ±20 inH2O to as low as ±0.25 inH2O (±50 to ±0.62 mbar). This unit has an accuracy of 0.05%FS with control stability better than 0.005%FS. All measurements can be made in differential or gauge pressures. |
ADT761A-D The Additel 761A-D also provides differential and gauge measurement which covers the range of -13.5 to 35 psi (-0.95 to 2.5 bar). This unit comes with a CP35 module (-13.5 to 35 psi) and one low range module of your choice ranging as low as ±10 inH2O (25 mbar). Each ADT761A-D can be preconfigured with the modules that fit your need to give you the best precision at the pressures you perform calibrations. |
ADT761A-500 The Additel 761A-500 will generate and control from vacuum pressures up to 500 psig (35 bar.g). Both gauge and absolute pressures can be realized due to a built-in barometer. Each unit comes with a CP500 module (-12.5 to 500 psig) for the high range and the low range can be preconfigured based on the variety of modules available down to 10 psig (0.7 bar.g). |
ADT761A-1K The Additel 761A-1K will quickly generate and control from vacuum pressures up to 1,000 psig (70 bar.g). Like the ADT761A-500, both gauge and absolute pressures can be realized due to a built-in barometer. Each unit comes with a CP1K module (-12.5 to 1,000 psig) for the high range and the low range can be preconfigured based on the variety of modules available down to 30 psig (2 bar.g). |
ADT761A-BP The Additel 761A-BP is designed for calibration of barometer sensors. With a range of 100 to 1200 hPa and an accuracy of 0.01%FS, this unit is ideal for calibration on the bench or in the field. |
ADT761A-APXR Precision Accuracy Options The Additel 761A series includes a precision accuracy option which provides an accuracy of 0.01%FS. This calibrator option includes a single non-removable sensor and can measure absolute and gauge pressures. Model configurations are available from 15 to 1,000 psig (1 to 70 bar.g). |
Applications
High Pressure Automatic Calibration
|
Pressure Gauge, Transmitter, or Switch Calibration
|
Task Menu
|
Specifications
|
|
|
|
ADDITEL DRY WELL TEMPERATURE CALIBRATOR
Sensor Thermocouple |
Additel 875-155 • Temperature range from -40°C to 155°C |
Sensor Thermocouple |
Additel 875-350 • Temperature range from 33°C to 350°C |
Sensor Thermocouple |
Additel 875-660 • Temperature range from 33°C to 660°C |
Sensor Thermocouple |
Additel 875PC-155 • Temperature range from -40°C to 155°C |
Sensor Thermocouple |
Additel 875PC-350 • Temperature range from 33°C to 350°C |
Sensor Thermocouple |
Additel 875PC-660 • Temperature range from 33°C to 660°C |
MULTIFUNCTION PROCESS CALIBRATORS
Additel 223A
|
|
Additel 222A
|
|
Additel 221A
|
|
Additel 220
|
|
Additel 210
|
|
Additel 209
|
PRESSURE CONTROLLERS
Additel 780
|
|
Additel 780S
|
|
Additel 770
|
|
Additel 770S
|
ADDITEL DIGITAL PRESSURE CALIBRATOR/GAUGES
APPLICATION • Precision pressure measurement • Field gauge calibration • Pressure safety valve (PSV) testing • Pressure regulator testing |
Additel 681
|
APPLICATION • Precision low pressure measurement |
Additel 681-DP
|
APPLICATION • Precision pressure measurement • Field gauge calibration • Pressure safety valve (PSV) testing • Pressure regulator testing |
Additel 680
|
APPLICATION • Precision pressure measurement • Field gauge calibration • Pressure safety valve (PSV) testing • Pressure regulator testing |
Additel 680W
|
APPLICATION • Gauge calibration • Precision pressure measurement • Pressure transmitters calibration • Pressure switch testing • Safety relief valve resting • Pressure regulator testing • Hydraulic leak testing |
Additel 672
|
APPLICATION • Gauge DP calibration • Pressure DP transmitters calibration |
Additel 672-DP
|